Instructions for Use Supplement C – Multi-Gas Sensor LeoLyzer
|
Rev. 02
21
8
Maintenance and repair
8.1
Device maintenance
Caution
Before performing any maintenance, ensure that the device and its
accessories are properly disinfected.
To ensure the proper functioning of the device during its service life, perform
regular maintenance.
The user of the device is responsible for proper maintenance. Maintenance work
is not covered by the warranty of the manufacturer or supplier.
Maintenance work does not require special skills or knowledge; it is not neces-
sary to disassemble and reassemble the electronic components for maintenance.
Even if the device is not used or temporarily stored, it must be checked for prop-
er functioning at least once per year.
8.1.1
Maintenance schedule
No. Task
Interval
Description
1.
Replace water trap
When neces-
sary
Section 8.1.2
2.
Replace rubber O-rings
Annually
Section 8.1.3
8.1.2
Replacing the water trap
• Disconnect the interface cable of the device from the intensive care ventilator.
• Disconnect the airway adapter from the breathing circuit.
• Disconnect the sampling line from the water trap.
• Remove the water trap by pressing the two tabs on the sides of the water
trap; then pull it carefully forward, away from the device.
• Install a new water trap: it must click into place.
8.1.3
Replacing the rubber O-rings
•
Use tweezers to remove the rubber O-rings from the fittings.
•
Again using tweezers, stretch the new rubber O-ring and place it on the fitting.
Apply the second rubber O-ring in the same way.
• Rubber O-rings are usually easy to stretch without tearing.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...