12
Instructions for Use Supplement D – Pulse Oximetry and LeoClac
|
Rev. 01
4.3
Function
• Use the XPOD SpO
2
sensor adapter to connect the SpO
2
sensor to the inten-
sive care ventilator (see chapter 4)
•
Apply the finger clip to the patient at a suitable site
• Select the SpO
2
function
• Set pulse wave averaging
• Adapt the SpO
2
and HR alarm limits to the clinical situation
• Start SpO
2
measurement (“On”).
Alarm limits
HR (pulse) max
Upper alarm limit for the measured heart rate
HR (pulse) min
Lower
alarm limit for the measured heart rate
SpO
2
max
Upper alarm limit for the measured
SpO
2
concentration
SpO
2
min
Lower alarm limit for the measured SpO
2
concentration
5
LeoClac (Closed-Loop Automatic Oxygen Control)
Prerequisite for the LeoClac function is the activated pulse oximetry function
(see chapter 4).
Warning
The LeoClac function is only available for the patient category
“Adults“.
5.1
Prerequisites
Further prerequisites on the configuration level:
• Select the LeoClac template
•
Configure the LeoClac function
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...