IF55 ROT CANopen®
Always replace the connection cap at the end of the operation. Take care in re-
connecting the internal connector. Tighten the screws
1
using a tightening
torque of approx. 2.5 Nm.
WARNING
You are required to check that the converter back flange and the connection
cap are at the same potential before replacing the connection cap!
Figure 5
4.2 SSI connection (Figure 4)
The converter is fitted with one M12 8-pin female connector to network the
IF55 gateway and the SSI encoder.
M12 8-pin
SSI connection
(frontal side)
A coding
female
Pin
Description
1
0Vdc power supply voltage
2
+Vdc power supply voltage
*
3
Clock IN +
4
Clock IN -
5
Data OUT +
6
Data OUT -
7 and 8
not connected
*
The power supply voltage level must be set through the POWER SUPPLY DIP
switch located inside the enclosure of the converter, see the “4.7 POWER
SUPPLY DIP switch (Figure 8)” section on page 29.
MAN IF55 ROT CB E 1.2.odt
4 - Electrical connections
26 of 88