99
Side Panels
1. Locate and remove the two Phillips screws along the front edge of each side
panel.
2. The side panels will hinge open once the screws are removed.
3. If side panel needs to be removed, position the panel so that it is almost closed.
Grasp the panel by the front and rear edges. Lift the panel up approximately 2
inches (5 cm), then pull it straight away from the end of the biosafety cabinet.
Product illustration
may not accurately
represent all models
Figure 12-2
Left
Side
Panel
Right Side Panel
Screw Location
Screw Location