83
17. Compare the inflow volume in CFM calculated in step #15 with the value in
column “Corrected Inflow Volume Range” based on Type and model shown in
Table 10-8.
18. If the Inflow Volume is not within the acceptance criteria range, adjust the
biosafety cabinet according to
Adjusting the Blower Speed
and/or
Adjusting the
Internal Damper
found in
Section 9: Calibration
.
Acceptance Criteria
Inflow Volume shall be in the range specified in Table 10-
8 column “Corrected Inflow
Volume Range”.