34
8: Configuration
This section provides instructions to access and understand the current configuration of
the biosafety cabinet, and make changes to the configuration.
Keypad button presses are shown as
[BLUE WITH BRACKETS]
. Menu screen
selections are shown as
green italics
.
Current Configuration
It is important to understand the current configuration of the biosafety cabinet for many
reasons, some of which include:
Verify model and type
Verify accessories installed
Troubleshooting
The Info Screen provides information on the current configuration of the biosafety
cabinets, and details the current blower speed(s). The Info Screen is not password
protected. To access this screen, follow these steps:
1. From the Main Menu, press
[MENU]
, using
[UP]
or
[DOWN]
select the
Tools
icon. Press
[OK/MUTE]
.
[MENU]
Tools
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...