77
4. Lower the sash until the skirt frame rests on the inside of the glass and remains
in place. A tall stand may be used to support the opposite end of the DIM if
desired.
5. If a sash alarm occurs, press
[OK/MUTE]
.
6. Install a blocking plate on either side of the DIM, and tape them securely on all
four sides. All air should enter through the DIM, seal gaps around blocking
plates completely.
7. Turn on the DIM, set it to read airflow volume continuously.
8. Take 5 readings over a 1 minute period. Average the results.
9. Compare results with
column “Inflow Volume Range” based on Type and model
shown in Table 10-7.
10. If the Inflow Volume is not within the acceptance criteria range, adjust the
biosafety cabinet according to
Adjusting the Blower Speed
and/or
Adjusting the
Internal Damper
found in
Section 9: Calibration
.
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...