179
7. If the password is entered correctly, you will be allowed
to select between altering the cabinet’s configuration,
or calibration. Press the
[DOWN]
button until the
Calibration
option is highlighted, then press
[OK/MUTE]
to start the calibration procedure.
IF THE UNIT IS CONNECTED TO AN EXHAUST
SYSTEM, YOU MUST STOP ALL AIRFLOW OVER
THE SENSOR DURING THE INITIAL ZEROING OF
THE SENSOR - DO THIS BY SHUTTING OFF THE
EXHAUST SYSTEM, OR BLOCKING THE SENSOR
INTAKE WITH TAPE.
8.
If the biosafety cabinet’s internal blower was on, it will automatically turn off.
Once all airflow from a remote (building) exhaust system has been removed (if a
Type A2 with canopy, Type B2, or Type C1 operating in B-mode), press
[OK/MUTE]
to continue.
9. The unit must sit for approximately 30 seconds with no
airflow over the sensor in order to establish the zero
point value.
10. After approximately 30 seconds, the cabinet blower(s)
will restart, and then the exhaust flow over the sensor
must be reestablished
– either open the exhaust valve,
or uncover the sensor, press
[OK/MUTE]
to continue.
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...