166
Kit Catalog Numbers
Installation Procedure
NOTE:
Before beginning this installation, Turn the unit on, press the
[MENU]
button, select
the
Tools
option, then
Information
. This will bring up the current settings screen. Record the
motor(s) average RPMs once the unit operation stabilizes. Type A2-12 inch and C1 models
will have two different (supply and exhaust) motor RPMs to record.
Exercise caution when removing or installing the exhaust filter cover, to prevent damaging
the Exhaust HEPA filter.
1. Turn off the Biosafety Cabinet.
Cabinet
Width (Feet)
Catalog
Number
3-ft
3889200
4-ft
3889201
5-ft
3889202
6-ft
3889203
Table 17-1
2. Locate and remove the four acorn
nuts and lockwashers that secure
each side of the exhaust cover
and the angle bracket to the top of
the unit (8 nuts & washers total).
3. Lift the cover and angles straight
up off of the top of the cabinet.
Keep the nuts and washers; you
may also want to retain the cover
and angles for the future, if the
cabinet will be disconnected from
the exhaust system.
Figure 17-3
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...