132
Changing the Optional UV Lamp
The UV Lamp is an optional feature and may not be found on all models.
For optimum performance, the UV lamp should be changed as indicated by the UV
lamp hourmeter found on the LCD Display in the
Settings
submenu.
The UV lamp and the work area of the cabinet must be thoroughly decontaminated
before removing the lamp.
1. Start the cabinet blower and let it operate for 5 minutes.
2. Raise the sash to its full open position.
3. Thoroughly surface decontaminate the UV lamp and the work area of the
cabinet.
4. Unplug the cabinet from all electrical power.
5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and out of
its sockets.
6. Install new lamp by reversing the removal procedure.
7. Reset the UV Hourmeter as follows:
In the
Settings
submenu enter
UV Parameters
, then select
Reset
, press
[OK/MUTE]
. The hour text will begin to flash, hold
[OK/MUTE]
for 3 seconds.
THE UV LAMP IN THIS PRODUCT CONTAINS MERCURY
Manage in accordance with local disposal laws. DO NOT place lamps in trash.
Dispose as a hazardous waste. For information regarding safe handling,
recycling and disposal, consult www.lamprecycle.org
CETTE LAMPE (UV) DANS CE PRODUIT CONTIENT DU MERCUE
Éliminez ou recyclez conformément aux lois applicables. Pour de l‘information
concernant des pratiques de manipulation sécuritaires et l’élimination sécuritaire
et le recyclage, veuillez consulter www.lamprecycle.org
Reset
UV
Parameters
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...