142
Resetting a Circuit Breaker
Should an over current situation arise during normal operation, the circuit breakers
located on the left side of the electronics module will trip, protecting the cabinet from
damage.
The biosafety cabinet offers internal outlets to power small devices inside the cabinet.
The internal outlets are protected by one circuit breaker (two circuit breakers on 230v
models), and the cabinet electronics are protected by a different circuit breaker (two
different circuit breakers on 230v models). This allows the cabinet to continue to
operate, providing protection to the product and user, should a device connected to an
internal outlet experience an over current situation.
If the internal outlets do not have power when the
Outlet
icon is displayed on the Home
Screen, reset the appropriate circuit breaker(s) as shown in Figure 13-2. The outlet
circuit breaker(s) are in the Rear Position(s).
If the cabinet does not have power when the System Reset Switch (Figure 13-2) is on,
reset the appropriate circuit breaker(s) as shown in Figure 13-2. The cabinet circuit
breaker(s) are in the Front Position(s).
The thrown circuit breaker will be easy to identify, as the white barrel will be extended
further than the other circuit breaker(s). To reset, simply push it in.
Outlet
Circuit Breaker
(Rear Position)
Cabinet Circuit Breaker
(Front Position)
Figure 13-2
System Reset Switch
Содержание Logic+ A2
Страница 148: ...148 Wiring Diagrams The wiring diagram is also located on the blower plenum cover behind the dress panel 100 115V...
Страница 149: ...149 208 240V...
Страница 151: ...151 Blower only will not start...
Страница 152: ...152 Lights only will not illuminate...
Страница 153: ...153 UV Light will not illuminate...
Страница 154: ...154 Airflow Alert activating...
Страница 155: ...155 Filter Life Gauge not at 100 when new...
Страница 156: ...156 Contamination in the work area...
Страница 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Страница 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...