27
Data Plate Information
The Data Plate provides the following important information for certifier use during on-
site certifications:
Nominal Airflows (Inflow and Downflow)
Downflow Grid Test Point Locations
Introduction Location for Test Aerosol
Secondary Inflow Methodology
NSF49 standard revision under which the biosafety cabinet is listed
Figure 6-2 shows a Type A2, 4-feet width Data Plate as an example to identify where
the required information is located on the Data Plate. All Labconco Data Plates are in
the same format.
Figure 6-2