6 | Function description
The [Management] tabulator with three dialog windows
6.12.3 The [device settings] dialog
In dialog [Device options], various pre-settings for the Rework System can be ac-
cessed.
Style
Device selection
Selected device
Rework
Archive
User administration
Profile management
Administration
HRSoft 2 options
Graph options
Device options
Firmware update
System board update file
Power board update file
Satellites update file
Device setup
Device transport position
Move to transport position
Firmware update
Start device setup
Select file
Select file
Select file
Temp. offset (120° C)
Channel TC 1
Reduce Z-axis velocity
Distance to reduce z-axis velocity
Send
Channel TC 2
Select file
Select file
Xmos update file:
Amtel update file:
Device selection
[Selected device]: select “HR 500“.
Reduce Z-axis velocity
– [Distance to reduce z-axis velocity]: For the last few millimetres before the con-
tact point, the pipette must slow down the contact movement. The 0 value cor-
responds to the PCB surface.
a) Enter the distance in millimetres at which contact movement must be slowed
down. Keep a higher distance for large components.
Temp. offset (120° C)
– [Channel TC 1] to [Channel TC 2]: to enter correction values if the temperature
readings of the thermocouples differ from the real temperatures.
a) Measure the thermocouples with an external calibration tool at 120°C.
b) Enter the values of the detected temperature differences (negative minus sign
for negative values) into input fields from [Channel TC 1] to [Channel TC 2] to
offset them.
Button [Send]
The temperature correction values were adopted and the Rework System will dis-
play the calibrated values of the temperature sensors.
Ersa GmbH
Operating_instructions HR 500 DE 3BA00246-01 | Rev. 1
88 / 124