399
10 .2 Error codes and countermeasures
Chapter 10
Error code
Displayed message
Problems/countermeasures
S30470
[AeroDR Stitching Unit name] Internal Error
Please power OFF/ON of AeroDR Stitching unit .
Restart the AeroDR Stitching System .
S30540
[AeroDR Stitching Unit name] MCB Error
Contact a service representative .
Contact Konica Minolta technical representatives .
S307**
[AeroDR Stitching Unit name] Internal Error
Please power OFF/ON of AeroDR Stitching unit .
Restart the AeroDR Stitching System .
S30FD0
[AeroDR Stitching Unit name] Internal Error
Contact a service representative .
Contact Konica Minolta technical representatives .
S30FE0
S30FF0
[AeroDR Stitching Unit name] Internal Error
Please power OFF/ON of AeroDR Stitching unit .
Restart the AeroDR Stitching System .
S40430
[AeroDR Stitching Unit name] Height detection failed
(lamp is off)
Please press the AeroDR Stitching unit RESET switch .
Repeat the height detection .
(1) Press the Reset switch on the AeroDR Stitching
X-Ray Auto-barrier Unit to cancel the error .
(2) Check that the exposure field lamp is set to on,
press the Set switch on the AeroDR Stitching
X-Ray Auto-barrier Unit, and re-execute height
detection .
(3) If the problem is not solved by the above method,
there is a possibility that the intensity of the expo
-
sure field lamp has changed. Change the bright
-
ness setting .
S40440
[AeroDR Stitching Unit name] Height detection
failed(cross mark mismatch)
Please press the AeroDR Stitching unit RESET switch .
Repeat the height detection .
(1) Press the Reset switch on the AeroDR Stitching
X-Ray Auto-barrier Unit to cancel the error .
(2) Check the installation position of the Collimator of
the X-ray device and the AeroDR Stitching X-Ray
Auto-barrier Unit is correct .
(3) If the problem is not solved by the above method,
there is a possibility
that the intensity of the expo
-
sure field lamp has changed. Change the intensity
settings and then implement height detection .
S40450
[AeroDR Stitching Unit name] Height detection
failed(other reason)
Please press the AeroDR Stitching unit RESET switch .
Repeat the height detection .
(1) Press the Reset switch on the AeroDR Stitching
X-Ray Auto-barrier Unit to cancel the error .
(2) Check the collimator installation position in the X-
ray device is correct .
(3) Execute height detection.
(4) If the problem is not solved by the above method,
there is a possibility that the intensity of the expo
-
sure field lamp has dropped. Reset the light sensor
sensitivity setting and then execute height detec
-
tion .
(5) If the problem is not solved with the light sensor
sensitivity setting, there is a possibility that the
intensity of the exposure field lamp has changed.
Change the brightness setting .
S40480
[AeroDR Stitching Unit name] Obstacle in the move
-
ment
Obstacle detected in the AeroDR Stitching unit move-
ment . Please press the RESET switch .
Check that there are no obstacles in the operating
range of the Stitching Mask of the AeroDR Stitching
X-Ray Auto-barrier Unit, and press the Reset switch
on the AeroDR Stitching X-Ray Auto-barrier Unit to
cancel the error .
S40490
[AeroDR Stitching Unit name] Switch Error
AeroDR stitching unit switch is in pressed state .
Confirm the switch not pressed state and switch the
power OFF/ON of AeroDR stitching unit .
Check that the Set switch of the AeroDR Stitching
X-Ray Auto-barrier Unit is not being continuously
pressed, and restart the AeroDR Stitching System .
S404A0
[AeroDR Stitching Unit name] Switch Error
AeroDR stitching unit switch is in pressed state .
Confirm the switch not pressed state and switch the
power OFF/ON of AeroDR stitching unit .
Check that the Image Area switch of the AeroDR
Stitching X-Ray Auto-barrier Unit is not being con-
tinuously pressed, and restart the AeroDR Stitching
System .
S404B0
[AeroDR Stitching Unit name] Switch Error
AeorDR stitching unit switch is in pressed state .
Confirm the switch not pressed state and switch the
power OFF/ON of AeroDR stitching unit .
Check that the Reset switch of the AeroDR Stitch-
ing X-Ray Auto-barrier Unit is not being continuously
pressed, and restart the AeroDR Stitching System .
Continued on the next page
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...