201
7.1 Exposure with AeroDR Portable RF Unit
Chapter 7
7 .1 .2 Exposure preparations
The exposure using the AeroDR Portable RF Unit is
performed with the following procedure . The following
procedure is for an example in which a wired connec
-
tion is in operation between this device and a higher
system .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When this device is being used in the AeroDR Por-
table RF Unit, charge it when a notification that the
battery level is low is displayed .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Register the DR Detector in this device that is
to be used in AeroDR Portable RF Unit.
• For operation of the DR Detector, refer to
"AeroDR Operation Manual" .
HINT
•••••••••••••••••••••••••••••••••••••
•
Confirm that the outputs of host, RIS-OUT and printer
are [ON] on the system monitor screen ([System Moni
-
tor] tab).
In case of the host
The device switches between [ON] and [OFF] every
time [Output] is pressed.
•
When using a RIS-OUT or printer, confirm the output is
set to [ON] using the same procedure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Import the examination information to be
used for the exposure using the portable
X-ray device .
•
Confirm that an Ethernet cable is connected to
this device .
•
Press [Refresh] on the list screen on this device.
•
The latest examination information is imported
from the higher system by pressing the [Refresh]
button .
HINT
•••••••••••••••••••••••••••••••••••••
•
To input examination information manually, change the
screen to the patient registration screen .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Remove the Ethernet cable that is con
-
nected to the DR Detector registration hub
from the portable X-ray device hub .
HINT
•••••••••••••••••••••••••••••••••••••
• When the AeroDR Portable RF Unit is connected to
this device through a wireless connection, remove the
Ethernet cable from this device .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
Unplug the power cable from the outlet
when charging this device .
5
Complete the exposure preparations .
• Confirm that the AeroDR Portable RF Unit is
shut down to complete the exposure and move
-
ment preparations .
HINT
•••••••••••••••••••••••••••••••••••••
• It is all right to keep power of this device and the DR
Detector on .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...