246
7.5 Manual stitching exposure
2
Press [OFF] or the button of the output
size to be changed on the Split Output
Size dialog box.
• The output size is changed .
HINT
•••••••••••••••••••••••••••••••••••••
• When output of split images is enabled, a frame ap-
pears on the image to indicate the area of the split im-
age that will be output .
Split output frame
Trimming frame
• The size of the trimming frame changes in accordance
with changes to the size of the image frame so that
images are automatically split into the selected section
size. Images can be split into a maximum of 3 sec
-
tions .
• If the split image is smaller than the trimming region,
the entire combined image may not completely fit
within the selected region .
•
If the configuration of the split images results in one
image, the split processing is not applied, and the split
output frame does not appear .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
7 .5 .2 Exposure
The manual stitching exposure method is as follows.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• For the stitchi
ng exposure images, make sure that the
combined parts are combined in the correct positions .
• If there is gap between combined images, manually
adjust the image position of the combined parts using
the viewer screen (for stitching combination position
adjustment) . It is also recommended that you arrange
a measure in the combined image part during expo
-
sure to confirm the combined parts easier.
•
In stitching exposure, images can be combined when
all the following conditions are met .
–
All exposures are done using DR Detectors (CR
cassettes) of the same size
–
All exposures are done using an DR Detector or a
CR cassette
–
Perform exposure after standardizing the read resolution
•
When performing the manual stitching exposure in
AeroDR 3 1417 HD, only "Standard" (200µm) can be
selected for "Resolution" .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
After transitioning from the initial screen to
the exposure body part selection screen, se-
lect the Exam . Tag for stitching exposure .
•
Select 2 or 3 Exam. Tags for stitching exposure.
•
The selected Exam. Tags are displayed on the
Exam. Tags display area.
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...