370
8.2 Dialog boxes/balloons and countermeasures
Displayed message
Problems/countermeasures
Study reserve to the console [Other device name] got failed
for the study [PATIENT_ID,PATIENT_NAME]. Not existing the
respective study . Please check the study status .
The target study has been deleted or has already been moved .
Check the target study status .
Study reserve to the console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. Patient information is
not matching, Please correct the patient information .
The patient information in the destination does not match the
patient information in the source .
Check the source and destination patient info . When the patient
name, date of birth, and sex are different, the patient is deter
-
mined to be someone else and it is not possible to move the
study .
Study reserve to the console [Other device name] got failed
for the study [PATIENT_ID,PATIENT_NAME].check the study
status or Study sharing status .
The target study does not satisfy sharing conditions . Check the
target study status . Target studies in the following statuses can-
not be moved .
•
The same study exists in the destination and is in progress
•
The same study exists in the destination and is complete
•
The same study exists in the destination and is scheduled for
moving
• The study in the source is in progress
• The study in the source is scheduled for moving to another
terminal
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. Check the network
status .
Communication between consoles failed .
Confirm that there is not a problem with the LAN connection
between the device and the console and establish the commu-
nications again .
If the problem persists, restart the device and try to establish
communications again .
If this does not solve the problem, contact Konica Minolta tech-
nical representatives .
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. Failed to access Im
-
age information of [Other device name]
Failed to move the image file due to unforeseen circumstances.
Move the target study again .
If the problem persists, restart the device and try to the target
study again .
If this does not solve the problem, contact Konica Minolta tech-
nical representatives .
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. Failed to access
study information of [Other device name]
Failed to move study info due to unforeseen circumstances .
Move the target study again .
If the problem persists, restart the device and try to move the
study again .
If this does not solve the problem, contact Konica Minolta tech-
nical representatives .
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. [Other device name]
DB access failed .
DB access failed due to unforeseen circumstances .
Move the target study again .
If the problem persists, restart the device and try to move the
study again . If this does not solve the problem, contact Konica
Minolta technical representatives .
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. [Other device name]
file access failed.
File access failed due to unforeseen circumstances .
Move the target study again .
If the problem persists, restart the device and try to move the
study again . If this does not solve the problem, contact Konica
Minolta technical representatives .
Study retreival from console [Other device name] got failed for
the study [PATIENT_ID,PATIENT_NAME]. Patient information is
not matching, Please correct the patient information .
The patient information in the destination does not match the
patient information in the source .
Check the source and destination patient info . When the patient
name, date of birth, and sex are different, the patient is deter
-
mined to be someone else and it is not possible to move the
study .
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...