9.1 Various problems and countermeasures
381
Chapter 9
9 .1 .5
Reacquisition of images
from the REGIUS
Reacquire the images when images cannot be re-
ceived from the REGIUS after the exposure due to
reasons such as communication errors, etc .
1
When the message "Image reception is
not completed . Check the network ." ap-
pears, press [OK] .
2
Please check the connection status with
REGIUS.
• A confirmation screen will be displayed . Press
[OK].
• The "Receiving the retransmitted images"
screen is displayed .
• It continues until the images are received suc-
cessfully .
• Images are reacquired and displayed .
HINT
•••••••••••••••••••••••••••••••••••••
• If you cancel via the "Receiving the retransmitted imag-
es" screen, a confirmation screen which says, "Are you
sure you want to cancel image reception? Restart the
reader in case of reception canceled ." will be displayed .
→ Press [OK] to cancel retransmission. The REGIUS
needs to be restarted .
→ Press [Cancel] to return to the "Receiving the retrans
-
mitted images" screen, and to continue receiving .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
9 .1 .6 Communications error
during image reception
When a fault occurs in network communications while
receiving images, operation sometimes becomes im-
possible .
When operation cannot be continued after about 5
minutes have elapsed, shut down this device and re-
start .
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...