8.2 Dialog boxes/balloons and countermeasures
Displayed message
Problems/countermeasures
Before study completion, please output or reject each Mammo
Image .
This is displayed when finishing an examination while there are
mammogram images that have not yet been output .
Output all mammogram images or reject them .
Below consoles are having more studies, but only partial stud-
ies are displayed .
Displayed when unable to acquire parts of a study after per-
forming multiple terminal study acquisition through study shar-
ing . Check the connection and reacquire the study .
Birth date is mandatory, Please enter birth date .
Displayed when mandatory fields have not been filled in.
Press [OK] to close the screen.
Calibration of [DR Detector name] have expired.
Please perform calibration urgently .
Do it now?
- If choose later:
Press the arrow buttons at the bottom of the screen .
And Press the calibration button in the toolbar that appears .
This is displayed when calibration of the DR Detector has ex
-
pired .
Perform calibration as soon as possible .
Calibration of [DR Detector name] will expire in [LimitDate] more
day(s) .
The calibration expiration date of the DR Detector is displayed.
Calibration of [DR Detector name] will expire in [LimitDate] more
day(s) .
Please perform calibration .
Do it now?
- If choose later:
Press the arrow buttons at the bottom of the screen .
And Press the calibration button in the toolbar that appears .
This is displayed when calibration of the DR Detector will be
expiring soon.
Perform calibration .
Can not manipulate the data in the deleted status .
The study cannot be opened as it has been deleted .
Cannot add exposures beyond [0].
Displayed when the number of scheduled exposures for the
same patient exceeds the maximum value. Further reservations
cannot be added .
Cannot add exposures beyond [MaxSize].
This is displayed when the maximum number of Exam. Tags
that can be registered has been exceeded.
Cannot capture images due to low battery power .
Charge the battery .
Displayed when the battery level for the DR Detector in use is
low . Charge the DR Detector .
Cannot edit the stitching position, because image stitching is
performed at other console . Please stitch the images again .
Displayed when trying to recombine an image that was com-
bined on another shared terminal . Combine the images again
on the original shared terminal .
Cannot get exposures beyond [0].
The Image Count for this patient has exceeded the upper limit.
No more studies can be acquired .
Cannot perform the calibration in the Technologist .
Please perform again in higher than super user privileges .
This is displayed when you log in as Technologist and try to
perform calibration .
Log in again using Super user or higher privileges .
Cannot perform the operation as selected study data save is in
progress .
Please perform the same operation after some time .
The selected study was displayed during image processing .
Wait until processing is finished and perform the operation
again .
Cannot perform the studies as different Patients studies got
selected .
Displayed when a different study is performed with a multi-
study . Reselect a study of the same patient .
Cannot perform the study as the study is in moving .
Displayed when the [Perform] button is pressed while a study is
being transmitted or the destination device is receiving a study .
Cancel sending the transmission or start the study with the
destination device .
Cannot perform this operation as image reading is inprogress
Displayed when attempting to shut down the system or delete
a job while reading an image . Shut down the system after the
image reading is finished.
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...