202
7.1 Exposure with AeroDR Portable RF Unit
7 .1 .3 Exposure
Prepare for exposure using the AeroDR Portable RF
Unit according to the following procedure .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
Always check the patient information and exposure
conditions before exposure.
• Do not disconnect the Ethernet cable while the data is
being transmitted .
• The status of communication with the host is displayed
on the [System] button. For detail, refer to
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
When you arrive at the exposure location, turn
on the power of the AeroDR Portable RF Unit.
• Turn on the power switch of the AeroDR Por-
table RF Unit, and check that the power LED
(green) is on .
• For startup operation of the AeroDR Portable RF
Unit, refer to "AeroDR Portable RF Unit Opera-
tion Manual" .
•
Confirm that the icon of the DR Detector to be
used in exposure is displayed on the application
bar .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
When the AP information channel is set to [Auto CH],
press [Auto CH] on the system monitor screen (por
-
table) and check the channel .
• Wireless connection may take several minutes after
the AeroDR Portable RF Unit is turned on .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Select the examination .
3
Press [Perform] .
•
The exposure screen is displayed.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD . Selecting the AeroDR 3 1417HD
icon allows the exposure time to be selected.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
•
When importing examination information from the
higher system and when exposure body part informa
-
tion is not imported, select the exposure body part.
When pressing [Perform] on the body part selection
screen, the exposure screen is displayed.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...