7.4 Exposure with Unitless
Chapter 7
4
Perform exposure .
•
Configure the portable X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
•
As soon as the exposed image is transferred,
the image is displayed in the image display area .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
When performing an exposure, stay away from the pa
-
tient as far as possible before pressing the exposure
switch .
• When the "Ready" is displayed, immediately perform
exposure.
• Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
•
If image acquisition started without performing ex
-
posure, this may result in incorrect detection due to
noise, etc . Perform reject (image deletion) and repeat
exposure.
•
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con
-
ditions and repeat exposure.
•
When exposure is done while "Ready" is not dis
-
played, have an interval of 1 minute or more before
repeating exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
5
Adjust the image .
•
After exposure, adjust the image as required.
6
Press [ ] .
•
The exposure is completed, and the display re
-
turns to the initial screen .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
When consolidating image adjustment, press [
] to
pause the examination and return to the initial screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• Image adjustment can be done on the portable de-
vice's [Image Proc] tab. However, if a more advanced
image adjustment is required, perform image adjust-
ment on the [Image Proc] tab or the viewer screen.
For details on the viewer screen, refer to "6 .7 Func-
tions of the viewer screen" .
• If you have to move to the imaging location, perform
the procedure again from step 1 after arriving to the
imaging location .
• It is all right to keep power of this device and the DR
Detector on .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is ON
1
After arriving to the imaging location, confirm
that the icon for the DR Detector used for im
-
aging appears in the application bar .
2
Select the examination .
3
Press [Perform] .
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...