7.3 Exposure with AeroDR Portable UF Unit
z
When the [AeroSync Button] is ON
1
When you arrive at the exposure location,
turn on the power of the AeroDR Portable
UF Unit .
• Turn on the power switch of the AeroDR Por-
table UF Unit and confirm that the (green) power
LED is ON .
• For startup operation of the AeroDR Portable UF
Unit, refer to "AeroDR Portable UF Unit Opera-
tion Manual" .
•
Confirm that the icon of the DR Detector to be
used in exposure is displayed on the application
bar .
IMPORTANT
•••••••••••••••••••••••••••••••••••••
•
When the AP information channel is set to [Auto CH],
press [Auto CH] on the system monitor screen (por
-
table) and check the channel .
• It may take several minutes for wireless connection to
start after turning on the AeroDR Portable UF Unit .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Select the examination .
3
Press [Perform] .
•
The exposure screen is displayed.
• A message "Waiting for permission" is displayed
in the image display area .
HINT
•••••••••••••••••••••••••••••••••••••
• If the information about the body part is not imported
while importing examination information from the high
-
er system, select the body part to be imaged . Then
press [Perform] on the body part selection screen to
display the exposure screen.
• When the AeroDR power management level is set to
[3], [4] or [5], "Waiting to start shooting" is displayed
instead of "Waiting for permission" .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
After positioning the patient, check the X-
ray detection level, then press [AeroSync] .
• If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
• Change the X-ray detection level as needed .
• After checking the X-ray detection level, press
[AeroSync].
• A message "Ready" is displayed in the mes-
sage/button area . The screen is locked .
• The screen is locked . To return to the previous
screen to change the position of the patient,
etc., press [Cancel].
Содержание CS-7
Страница 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Страница 2: ......
Страница 8: ...8 ...
Страница 9: ...9 Introduction ...
Страница 18: ...18 ...
Страница 30: ...30 ...
Страница 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Страница 34: ...34 ...
Страница 74: ...74 ...
Страница 75: ...75 Chapter 4 General Operations This chapter describes general operation methods of this device ...
Страница 84: ...84 ...
Страница 120: ...120 ...
Страница 121: ...121 Chapter 6 Functions of each screen This chapter describes the functions of each screen of this device ...
Страница 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Страница 282: ...282 7 9 SIGMA stitching exposure 2 Press Return when image adjustment is completed The exposure screen is displayed ...
Страница 353: ...353 Chapter 8 Displayed Messages This chapter describes the displayed messages and the countermeasures ...
Страница 375: ...375 Chapter 9 Troubleshooting This chapter describes the countermeasures when the problems occur ...
Страница 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Страница 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Страница 416: ...416 ...
Страница 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Страница 422: ...422 ...
Страница 423: ......
Страница 424: ...A47FBA01EN15 2017 01 12 JD ...