46
PWX
CONFIG Settings
CF02 Power-on status parameter
Sets the condition that the PWX series will be in when the power is turned on. This setting is
invalid when you are using an external contact to turn the output off.
This parameter is applied when you turn the POWER switch off and on.
You can only specify this parameter when the PWX series is being used as a master unit
.
CF03 Method for clearing OHP, FAN, and AC-FAIL alarms
Sets the method for clearing alarms when overheat protection (OHP), fan failure protection
(FAN), or low AC input protection (AC-FAIL) has been activated. You cannot set this parame-
ter separately for each protection function.
When you select “Auto,” if the overheat protection function (OHP) has been activated, output
will turn on again automatically when the internal temperature drops to a normal level. How-
ever, if you do not fix the problem that caused the alarm to occur, the alarm will occur again,
which will force you to clear the alarm again. The overheat protection devices have long ser-
vice lives (approximately 100000 uses), but if they reach the end of their service lives, they
will stop operating. To maintain the performance of the PWX series for a long time, we
request that you fully understand the Auto feature and only use it when necessary.
CF04 Memory content display parameter
Sets whether to display the saved contents in the preset memory before recalling them.
You can use this CONFIG parameter if you forget what settings have been saved to a preset
memory entry or if you want to view the saved settings.
You can only specify this parameter when the PWX series is being used as a master unit
.
CF05 CC control using an Vext or Rext
Sets whether constant current will be controlled by an external voltage (Vext) or an external
resistance (Rext) (the J1 connector).
This is set to “oFF” when the panel settings are reset (CF00) and when a preset memory
entry is recalled.
You can only specify this parameter when the PWX series is being used as a master unit
.
See
Value
Parameter description
SAFE
The settings that were in use immediately before the POWER switch was turned off
are used. The output is off (factory default setting).
Auto
The settings that were in use immediately before the POWER switch was turned off
are used. This includes whether output was on or off.
Forc
The settings that were in use immediately before the POWER switch was turned off
are used. The output is turned on.
See
See
Value
Parameter description
SAFE
After the problem that caused an alarm to occur is fixed, the output is not turned on
automatically (factory default setting).
Auto
After the problem that caused an alarm to occur is fixed, the output is turned on
automatically.
See
Value
Parameter description
oFF
The saved contents of preset memory will be recalled without displaying them.
on
The saved contents of preset memory will be displayed and then recalled after con-
firmation (factory default setting).
See
Value
Parameter description
oFF
Control will not be performed (factory default setting).
on
Control will be performed.
Содержание PWX1500H
Страница 10: ...10 PWX This page is intentionally blank...
Страница 26: ...26 PWX This page is intentionally blank...
Страница 88: ...88 PWX This page is intentionally blank...
Страница 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Страница 92: ...92 PWX This page is intentionally blank...
Страница 93: ...Specifications This chapter contains the specifications and gives the dimensions of the PWX series...
Страница 120: ...120 PWX This page is intentionally blank...
Страница 121: ...Appendix A Options B Troubleshooting...
Страница 128: ...128 PWX This page is intentionally blank...