50
PWX
CONFIG Settings
CF19 Current and power display on slave units for master-slave
parallel operation
Sets whether the current or power on slave units is displayed during master-slave parallel
operation. You can display and specify this parameter only when the PWX series is being
used as a slave unit.
If this parameter is set so that the current or power on slave units is not displayed, the amme-
ter displays
”
----.
”
CF20 LAN interface parameter
Sets whether the LAN interface will be used.
By setting it to “oFF,” you can disable the LAN interface function even when a LAN cable is
connected. You will also not be able to set the LAN interface parameters.
Even when this is set to “on,” if the command language is set to legacy (CF21: LGCy), you
cannot use the LAN interface.
This parameter is applied when you turn the POWER switch off and on.
CF21 Command language parameter
Sets the command language that is used in remote control messages.
This parameter is applied when you turn the POWER switch off and on.
CF22 Emulation parameter during remote control
Sets the emulation that is used during remote control. If you have set the command language
to legacy languages, you can only select “Gen” or “PAG.”
This parameter is applied when you turn the POWER switch off and on.
Value
Parameter description
oFF
The current or power of slave units is not displayed (factory default setting).
on
The current or power of slave units is displayed.
Value
Parameter description
oFF
The LAN interface will be not used
on
The LAN interface will be used (factory default setting).
Value
Parameter description
SCPi
The SCPI language will be used (factory default setting).
LGCy
Legacy languages will be used.
Value
Parameter description
nonE
Remote control is performed using the PWX series commands made by Kikusui
(factory default setting).
5700
Remote control is performed using the N5700/N8700 commands made by Agilent
Technologies.
GEn
Remote control is performed using the Genesys series commands made by TDK-
Lambda.
PAG
Remote control is performed using the PAG series commands made by Kikusui.
dCS
Remote control is performed using the DCS series commands made by Sorensen.
Содержание PWX1500H
Страница 10: ...10 PWX This page is intentionally blank...
Страница 26: ...26 PWX This page is intentionally blank...
Страница 88: ...88 PWX This page is intentionally blank...
Страница 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Страница 92: ...92 PWX This page is intentionally blank...
Страница 93: ...Specifications This chapter contains the specifications and gives the dimensions of the PWX series...
Страница 120: ...120 PWX This page is intentionally blank...
Страница 121: ...Appendix A Options B Troubleshooting...
Страница 128: ...128 PWX This page is intentionally blank...