60
PWX
CONFIG settings
Param.
number
CONFIG parameter
Parameter description
CF01
Master-slave parallel operation parameter
1 (independent operation)
CF02
1
Power-on status parameter
SAFE (panel settings that were in use the last time the
power was turned off)
CF03
Method for clearing OHP, FAN, and AC-FAIL
alarms
SAFE (do not clear the alarm automatically)
CF04
Memory content display parameter
ON (displayed)
CF05
CC control using an Vext or Rext
OFF
CF06
CV control using an Vext or Rext
OFF
CF07
CV/CC control range parameter
LO (0 V to 5 V)
CF08
Range during voltage and current monitoring
LO (0 V to 5 V)
CF09
External control parameter for turning output
on and off
OFF
CF10
External control logic parameter for turning
output on and off
LO (a LOW signal turns output on)
2
CF11
Bleeder on/off parameter
ON (bleeder circuit is on)
CF12
Output-on startup state parameter
CV (CV is prioritized)
CF13
Detection time of OCP activation
0 ms
CF14
Current setting limit
ON (the setting is limited)
CF15
Voltage setting limit
ON (the setting is limited)
CF16
Panel display brightness setting
7
CF17
Communication monitor timer setting
OFF (Communication monitoring timer off)
CF18
3
Internal resistance setting
OFF (Variable internal resistance feature is not used)
CF19
4
Display setting on slave units for master-slave
parallel operation
OFF (not displayed on slave units)
CF20
LAN interface parameter
LAN (The LAN interface will be used)
CF21
Command language parameter
SCPI
CF22
Emulation parameter during remote control
NONE (PWX series commands)
CF23
SCPI communication error display parameter
OFF (not displayed)
CF24
Multichannel (VMCB) domain number
0 (the multichannel function is not used)
CF25
Multichannel (VMCB) channel number
0 (the PWX is used as a master unit on the multichan-
nel network)
CF41
IP address assignment method
(DHCP: ON, AUTO IP: ON, MANUAL IP: OFF)
CF47
SCPI-RAW protocol TCP port number
5025
CF52
RS232C baud rate parameter
19.2 (kbit/s)
1
You can only specify this parameter when the PWX series is being used as a master unit.
2
0 V to 0.5 V or shorted.
3
You can only specify this parameter only when the variable internal resistance feature (factory option) is
installed.
4
You can only specify this parameter only when the PMX series is being used as a slave unit.
Содержание PWX1500H
Страница 10: ...10 PWX This page is intentionally blank...
Страница 26: ...26 PWX This page is intentionally blank...
Страница 88: ...88 PWX This page is intentionally blank...
Страница 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Страница 92: ...92 PWX This page is intentionally blank...
Страница 93: ...Specifications This chapter contains the specifications and gives the dimensions of the PWX series...
Страница 120: ...120 PWX This page is intentionally blank...
Страница 121: ...Appendix A Options B Troubleshooting...
Страница 128: ...128 PWX This page is intentionally blank...