PWX
CONFIG Settings
CF11 Bleeder on or off
Turns the bleeder circuit on and off. You can turn the bleeder circuit off when you don’t want
current to be sinked by the internal bleeder circuit, such as when you are charging a battery.
Even when the output terminal is off (open) or when the set voltage is 0 V, if you turn the
bleeder off, a voltage as large as a few hundred millivolts may appear at the output terminal.
Depending on whether the breeder circuit is on or off, the sink current and the fall time of out-
put voltage when an external voltage sources connected will vary.
When using master-slave parallel operation or series operation, use the same settings on all
connected units.
CF12 Output-on startup state parameter
Sets the operation mode to be prioritized when the output is turned on.
When used as a slave device, CC is automatically prioritized. To use the PWX series as a
standalone unit or master unit after using it as a slave unit, set the operation mode that you
want to use.
If you do not set the operation mode, CC will be prioritized.
You can only specify this parameter when the PWX series is being used as a master unit
.
CF13 Detection time of OCP activation
Set the time when the overcurrent protection (OCP) is activated after an overcurrent is
detected. Set this in 100 ms steps.
CF14 Current setting limit
Set whether to limit the set output current so that its value does not exceed the set overcur-
rent protection (approximately 95 % of the OCP trip point). When the current setting limit is
set to “oFF,” if you set the current setting limit to “on” when the set OCP is lower than the set
current, the current value will be maintained, and the set OCP will be changed to 105 % of the
current value. The parameter will also be changed if the output is on.
This is set to “on” when the panel settings are reset (CF00) and when a preset memory entry
is recalled.
You c
an only specify this parameter when the PWX series is being used as a master unit
.
See
Value
Parameter description
oFF
The bleeder circuit is turned off.
on
The bleeder circuit is turned on (factory default setting).
See
Value
Parameter description
CC
CC (constant current) is prioritized.
CV
CV (constant voltage) is prioritized (factory default setting).
See
Value
Parameter description
0
0 ms (factory default setting)
100 to 2000
100 ms to 2000 ms
See
Value
Parameter description
oFF
The setting is not limited.
on
The setting is limited (factory default setting).
Содержание PWX1500H
Страница 10: ...10 PWX This page is intentionally blank...
Страница 26: ...26 PWX This page is intentionally blank...
Страница 88: ...88 PWX This page is intentionally blank...
Страница 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Страница 92: ...92 PWX This page is intentionally blank...
Страница 93: ...Specifications This chapter contains the specifications and gives the dimensions of the PWX series...
Страница 120: ...120 PWX This page is intentionally blank...
Страница 121: ...Appendix A Options B Troubleshooting...
Страница 128: ...128 PWX This page is intentionally blank...