88
Keysight N4372E Lightwave Component Analyzer User’s Guide
5
User Interface Reference
RF Path Deembedding
Enable
Select if you want the LCA to compensate for the RF connections to the
DUT.
Use the results of a measurement of the RF path to characterize the
source or receiver RF path.
Figure 22
Graphical user interface of the LCA software with enabled RF path
deembedding
• For Single-Ended measurements, Src 1 corresponds to the path to
network analyzer Port 1, and Rcv 1 corresponds to the path to Port 2 on
a two-port network analyzer or, respectively, to Port 4 on a four-port
network analyzer.
• For Differential measurements, Src 1 corresponds to the path to Port 2
(on the network analyzer), Src 2 corresponds to the path to Port 3 (on
the network analyzer), Rcv 1 corresponds to the path to Port 2 (on the
network analyzer), Rcv 2 corresponds to the path to Port 3 (on the
network analyzer).
The results need to be in the form of a 2-port S-parameter data file (also
known as Touchstone .s2p data files). All the S-Parameters in the supplied
file are used.
Please note that 4-port S-parameter files (e.g. for differential probes) are
currently not supported.