Keysight N4372E Lightwave Component Analyzer User’s Guide
107
Definition of Terms
6
Figure 32
Balanced Measurement types for opto-electronic components
Calibration Concept
The key to making accurate EO, OE, or OO measurements is calibrated
instrumentation.
There are a number of calibration stages involved.
The first stage is the electrical calibration of the network analyzer.
This calibration establishes the electrical calibration reference planes for
the measurements to be at the RF output port of the frequency extender
heads. Whenever there are additional electrical adapters or elements
required for connecting the device under test these electrical paths can be
accounted for by "RF path deeembedding" (see
The second stage is the calibration of the optical test set. This is handled
as a “Fixture” by the network analyzer.
For more information on Fixtures, please refer to the documentation of the
network analyzer.