Keysight N4372E Lightwave Component Analyzer User’s Guide
43
Performance Quick Check
3
6 Set the following parameters:
•
If you have option 102 (Dual wavelength 1310 nm, 1550 nm), select
the appropriate wavelength for your DUTs.
•
Laser power: +3 dBm
•
Optical input: Input 1
•
User Calibration: None
•
Forward RF Power: 0 dBm
7 Click
, then
to set the output power.
8 Press
to start optical power meter application.
9 Read the optical power from the power meter application.
The readout should be +3 dBm ± 0. 5 dB.
10 Press the [Macro/Local] utility button until the Macro 1 group is
highlighted and you see
.
11 Press
to start the LCA measurement setup.