Performance Quick Check
3
8 Set the following parameters:
•
If you have option 102 (Dual wavelength 1310 nm, 1550 nm), select
the appropriate wavelength for your DUTs.
•
Laser power: +3 dBm
•
User Calibration: None
•
Forward RF Power: 0 dBm
9 Click
, then
to set the output power.
10 Right click the measurement trace, and select “Autoscale All” in the
context menu.
At the end of the measurement, there should be a flat trace.
An ideal trace might look like this: