96
Keysight N4372E Lightwave Component Analyzer User’s Guide
6
Definition of Terms
EO measurements (lasers, modulators)
The measurement of an EO transducer is a combination of input
modulating current and output optical modulation power. Slope
responsivity is used to describe how a change in input current produces a
change in optical power. Graphically this is shown in the following figure.
An LCA measures input modulating current and output modulation power
and displays the ratio of the two in Watts/Amp, either linearly or in
decibels.