Keysight N4372E Lightwave Component Analyzer User’s Guide
81
User Interface Reference
5
Second saving method
After initializing with the User Calibration mode set to “Measure”, you are
asked if you want to save the user calibration data and/or the LCA Setup.
Check “Also save the LCA Setup” if you want to save the LCA setup. Then
choose between “Load” and “Measure” to specify the user calibration
mode to save with the setup.
• If you selected “Load”, the “No” button is disabled to force you to store
the measured user calibration data.
After you press “Yes”, you are asked to specify the filename for the user
calibration data and the filename of the LCA setup.
Only the filename of the user calibration data is saved with the setup.
• If you selected “Measure”, it is also possible to press the “No” button. In
this case only the LCA setup is saved.
If you press “Yes”, you are asked for the filename of the LCA setup. The
name of the electrical calibration set just used for this user calibration
measurement will be saved in the LCA setup.
Before the LCA saves the setup, it asks you whether you want to save the
network analyzer state with your LCA setup.
Loading setups
Click
to retrieve an existing setup.
• After loading a setup with user calibration mode set to “Measure”, the
electrical calibration set specified in this setup is preselected in the
optical connection dialog shown during OE or OO initialization.
This is different to the behavior when doing OE or OO initialization
without loading a setup.
For more details, see
Measuring Opto-Electrical (OE) devices
Содержание N4372E
Страница 1: ...User s Guide Keysight N4372E Lightwave Component Analyzer ...
Страница 8: ......
Страница 22: ......
Страница 30: ...30 Keysight N4372E Lightwave Component Analyzer User s Guide 2 Setting Up the Instrument ...
Страница 92: ......
Страница 130: ......
Страница 136: ......
Страница 137: ...Keysight N4372E Lightwave Component Analyzer User s Guide 10 Definition of Terms Definition of Terms 138 ...