46
Keysight N4372E Lightwave Component Analyzer User’s Guide
3
Performance Quick Check
Verifying the Electrical Connection and Electro-optical Operation
The following tests ensure that the internal transmitter and receiver of the
Lightwave Component Analyzer operate as expected.
1 Follow the steps in the sections
analyzer settings and electrical calibration
2 On the network analyzer make the following settings:
•
IF bandwidth: 300 Hz
•
Select “Reduce IF BW at Low Frequencies”.
3 If you have option -311 and -021:
Connect the two straight-angled optical patch cords (N4373-87907) to
the optical output of the LCA Optical Transmitter and input of the LCA
Optical Receiver. Note that the angled connector must be attached to
the optical test set.
Use the FC/PC adapter (1005-0256) to connect the straight ends of the
optical patch cords.
If you have option -311 and -022:
Connect the angled-angled optical patch cords (N4373-87906) to the
optical output of the LCA Optical Transmitter and input of the LCA
Optical Receiver.
Use the FC/APC through adapter (1005-1027) to connect the ends of
the optical patch cords.
4 Connect the two supplied flexible test port cables to the RF ports of
the Lightwave Component Analyzer test set.
5 Press the [Macro/Local] utility button until the Macro 1 group is
highlighted and you see
.
6 Press
to start the LCA measurement setup.
7 Select the OO measurement tab.
Содержание N4372E
Страница 1: ...User s Guide Keysight N4372E Lightwave Component Analyzer ...
Страница 8: ......
Страница 22: ......
Страница 30: ...30 Keysight N4372E Lightwave Component Analyzer User s Guide 2 Setting Up the Instrument ...
Страница 92: ......
Страница 130: ......
Страница 136: ......
Страница 137: ...Keysight N4372E Lightwave Component Analyzer User s Guide 10 Definition of Terms Definition of Terms 138 ...