KLP 070312
4-7
increase and the “–“to decrease the output voltage until the DVM reads as close as possible
to minimum rated POSITIVE voltage.
FIGURE 4-2. VOLTAGE CALIBRATION WINDOW
2. Click the
Press to ACCEPT
button to accept the value. The status display then reads
VMAX
. Monitor
DVM and click the “+” button to increase and the “–“ to decrease the output
voltage until the DVM reads as close as possible to maximum rated voltage.
3. Click the
Press to ACCEPT
button to accept the value; the status display now reads
OUTV
.
At the ANALOG I/O PORT J2, connect the DVM (M– to pin 13, M+ to pin 6) and the d-c
power source (V– to pin 9, V+ to pin 3); set the d-c source output to 10.00V. Click the “+” but-
ton to increase and the “–“ button to decrease the output voltage until the DVM reads as
close as possible to 10V.
4. Click the
Press to ACCEPT
button to accept the value. The status display will not change.
At the calibration window, click the
CURRENT
Button to proceed to current calibration (see
PAR. 4.4.2)
.
4.4.2
CURRENT CALIBRATION (VXI
plug&play
DEMO)
During current calibration, the current, current readback and analog current readback are cali-
brated.
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......