3-32
KLP070312
Create the List.
Click
LIST
to open the List dialog box (Figure 3-12). Each location defines the
values for one step: voltage and current, and a dwell time duration (between 0.5 and 655.36
seconds).
NOTES: 1. Dwell times down to 0.01 second can be obtained using SCPI commands (via RS
232 or GPIB interfaces) to generate the list (see KLP Developer’s Guide).
2. The state of the internal relay can be stored using the VXI
plug&play
driver or SCPI
LIST commands (see KLP Developer’s Guide), but can not be controlled from the
web interface
Click
Add to List
button to add a step (the information displayed in the Voltage, Current and
Dwell boxes) to the List steps shown to the right.
FIGURE 3-12. LIST DIALOG BOX
Clearing the List.
Once a LIST has been started, the individual steps can not be changed; the
only option is to delete the LIST by pressing the
Clear List
button to start over. In addition, per-
forming any of the following operations will also clear the list: changing overcurrent or overvolt-
age PROTECTION settings, unit RESET (clicking *RST), changing VIRTUAL MODEL settings
and exiting the OPERATE INSTRUMENT page.
Create Voltage or Current Ramps.
In addition to adding single steps, the user may add multi-
ple steps to produce voltage or current ramps. Clicking the
Voltage Ramp
or
Current Ramp
button opens similar dialog boxes (see Figure 3-13). Enter the starting and ending ramp values,
the number of intermediate steps (including start and end), the total time (0.5 to 655.36 sec-
onds) to go from the start to the end of the ramp, and a constant current (for voltage ramp) or
voltage (for current ramp) for each step of the ramp. Clicking the
Create Ramp
button adds the
steps to the list as the firmware calculates the intermediate voltage values and dwell times
required to produce the desired ramp (calculated dwell times for each step must be greater than
0.5 second). Ramps with many total steps may take a few seconds to process; the percentage
complete is indicated below the
Create Ramp
button.
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......