1-6
KLP 070312
PROGRAMMING CHARACTERISTICS - ANALOG
Analog remote control
selection Activate with jumper at analog program-
ming connector
Recognized during power up.
isolation Safety Extra Low Voltage (SELV)
Analog Input Update Rate
2Hz (0.5 Second) Applies to program-
ming by voltage/ resistance and read-
back specifications.
Analog input voltage digitized (12-bit
resolution), optically isolated, then pro-
cessed by digital section.
Programming by voltage
voltage 0-10V
See PAR. 4.3.3 to reduce Full Scale
voltage.
current 0-10V
Min. programming limit; see Table 1-2.
See PAR. 4.3.3 to reduce Full Scale
voltage.
Programming by resistance
voltage 0-10K ohms
See PAR. 4.3.3 to reduce Full Scale
resistance.
current 0-10K ohms
Min. programming limit; see Table 1-2.
See PAR. 4.3.3 to reduce Full Scale
resistance.
Readback
0-10V proportional signal
Proportional to analog control volt-
age/resistance.
Remote inhibit
TTL compatible
Dual polarity; see PAR. 3.2.16.3.1.
Composite status flag
Isolated form C contacts
Programmable functions (PAR.
3.2.16.1).
PROGRAMMING CHARACTERISTICS - DIGITAL
Supported Interfaces
standard models GPIB and RS 232
SCPI command set for GPIB and RS
232 (see KLP Developer’s Guide).
E-Series models GPIB and LAN
Support SCPI command set (see KLP
Developer’s Guide) for GPIB and LAN;
Support four interfaces for LAN:
1. Web interface, port 80 (LXI Class C,
Version 1.2
2. SCPI Telnet, port 5024
3. SCPI Sockets, port 5025
4. VXI 11, port 1024
GPIB
GPIB address range: 1 to 30
Factory default is 6.
RS 232
standard models Baud Rate range: 2400, 4800, 9600,
19,200 or 38,400
Factory default is 38,400.
Digital remote control
isolation Safety Extra Low Voltage (SELV)
format compatible with SCPI protocols
W98 SE and later operating systems.
Programming resolution
0.024% of Emax and Imax
Programming accuracy
0.05% of Emax and Imax
Readback resolution
0.024% of Emax and Imax
Readback accuracy
Either 0.1% of Emax and Imax, or front
panel voltmeter or ammeter display res-
olution, whichever is greater.(1)
±1 LSB error
Status reporting
OVP, OCP, OTP, output lead fault
(OLF), fan failure, power loss
See PAR. 1.5.4 and 1.5.9.
(1) Display resolution is determined by the number of significant digits shown, relative to the decimal point. Resolution is 10-X where
X is the number of digits to the right of the decimal point. E.g for display of 10.00V, resolution is 10-2 or 0.01 or 10mV; for display of
150.0A, resolution is 10-1 or 0.1 or 100mA.
TABLE 1-3. KLP SPECIFICATIONS (Continued)
SPECIFICATION
RATING/DESCRIPTION
CONDITION/COMMENT
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......