3-10
KLP070312
4. Rotate the
CURRENT
control to change the GPIB address, then tap the
DC OUTPUT
switch
to accept the displayed GPIB address. The display briefly reads
DONE
, then returns to
ADDR
.
5. Rotate the
VOLTAGE
control until display reads
EXIT
, then tap the
DC OUTPUT
switch to
exit
3.2.12
CHANGING RS 232 BAUD RATE [STANDARD MODELS ONLY]
The factory default RS 232 baud rate is 38400. To change the baud rate from the front panel
proceed as follows:
1. If the status display shows
SET
, tap either the
CURRENT
or
VOLTAGE
controls to take the
unit out of setpoint mode (status display goes from
SET
to blank)
2. Using a thin tool (e.g., a paper clip), press the
FUNCTION
switch once, then rotate either the
CURRENT
or
VOLTAGE
control until the display reads
BAUD
; the active baud rate 1000 x
the value visible in the
DC AMPERES
display (e.g., 38400 is displayed as 38.4). To change
the baud rate, proceed to step 3.
3. Tap the
DC OUTPUT
switch. The baud rate visible in the
DC AMPERES
display begins
blinking.
4. Rotate the
CURRENT
control to change the baud rate, then tap the
DC OUTPUT
switch to
accept the displayed baud rate. The display briefly reads
DONE
, then returns to
BAUD
.
5. Rotate the
VOLTAGE
control until display reads
EXIT
, then tap the
DC OUTPUT
switch to
exit
3.2.13
SETTING UP MASTER/SLAVE CONFIGURATIONS [STANDARD MODELS ONLY]
Both units of a master/slave pair must be identical models. After connecting the RS 232 ports
(see PAR. 2.7.8), proceed as follows to configure the individual units. First configure the slave
(see PAR 3.2.13.1), then the master (see PAR 3.2.13.2).
3.2.13.1 CONFIGURE POWER SUPPLY AS SLAVE [STANDARD MODELS ONLY]
1. Turn power on. If the unit comes up in setpoint mode (see PAR. 3.2.6), tap either the
CUR-
RENT
or
VOLTAGE
control to take the unit out of setpoint mode (status display goes from
SET
to blank).
2. Using a thin tool (e.g., a paper clip), press the
FUNCTION
switch once, then rotate either the
CURRENT
or
VOLTAGE
control until the display reads
M/S
.
3. Tap the
DC OUTPUT
switch.
• The status display reads either
SERI
(series) or
PARA
(parallel).
• The
DC VOLTS
display shows the number assigned to the unit (1 = master, 2 or above =
slave).
• The
DC AMPERES
display shows the maximum number of units in the configuration.
• The factory default is standalone operation (
DC VOLTS
and
DC AMPERES
show
1
).
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......