KLP070312
3-13
3.2.13.4 RESTORING THE MASTER/SLAVE CONFIGURATION TO STANDALONE OPERATION
[STANDARD MODELS ONLY]
First configure the master, then the slave for standalone operation.
1. If the units are in setpoint mode (see PAR. 3.2.6), tap either the
CURRENT
or
VOLTAGE
controls to take the units out of setpoint mode (status display goes from
SET
to blank).
2. Using a thin tool (e.g., a paper clip), press the
FUNCTION
switch of the unit to be restored to
standalone operation, then rotate either the
CURRENT
or
VOLTAGE
control until the display
reads
M/S
.
3. Tap the
DC OUTPUT
switch once.
• The status display reads either
SERI
(series) or
PARA
(parallel).
• The
DC VOLTS
display shows the number assigned to the unit (1 = master). The
DC
AMPERES
display shows the maximum number of units in the configuration.
4. To change the configuration tap the
DC OUTPUT
switch once. The
DC VOLTS
and
DC
AMPERES
start blinking.
• Rotate the
CURRENT
control until
DC AMPERES
display shows
1
.
• Rotate the
VOLTAGE
control until
DC VOLTS
display shows
1
.
5. Press
DC OUTPUT
to accept the displayed settings. The unit briefly flashes
WAIT
, then
restarts with the new settings; press
FUNCTION
to abort without changing the settings.
NOTE: Once the master is configured for standalone operation, the slave status display will
flash
E241
(hardware error).
6. Repeat steps 2 through 5 to restore the slave unit to standalone operation.
3.2.14
LAN INTERFACE CONFIGURATION [E-SERIES MODELS ONLY]
Using a thin tool (e.g. end of paper clip), press
FUNCTION
switch once, then rotate either
VOLTAGE
or
CURRENT
control until status display reads
LAN
. Press
DC OUTPUT
switch to
enter the LAN Interface configuration menu. Rotate either
VOLTAGE
or
CURRENT
control to
see the available functions.
•
LRST
- LAN Reset. To restore the LAN interface configuration to the factory default, tap
DC OUTPUT
switch once, then rotate either
VOLTAGE
or
CURRENT
control to show
YES
in the
DC VOLTS
display, then tap
DC OUTPUT
switch again. The factory default
settings listed in PAR. 3.3.1.3 are restored upon exiting the Function menu.
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......