4-6
KLP 070312
The main panel status display shows calibration, and the voltage and current meters are not
visible.
FIGURE 4-1. CALIBRATION WINDOW
7. Use the computer keyboard to enter the password in the text box displaying the word PASS-
WORD, then press the ENTER key.
8. If password is correct, the status display on the VXI main panel reads CAL and the buttons
on Calibration Screen (Figure 4-1) are enabled (highlighted). If the password is incorrect, the
password text box displays “WRONG #”. Repeat step 7 to enter the correct password, or
click the
Calibrate Close
button on the VXI main panel to exit calibration.
NOTE: Calibrations may be done in any order.
9. Proceed to PAR. 4.4.1 for voltage calibration, PAR 4.4.2 for current calibration., PAR. 4.4.3
for external calibration or PAR. 4.4.4 to exit calibration
4.4.1
VOLTAGE CALIBRATION (VXI
plug&play
DRIVER DEMO)
During voltage calibration, the voltage, voltage readback and remote analog voltage readback
are calibrated.
1. Click the
VOLTAGE
button on the Calibration window to start voltage calibration: the Voltage
Calibration Window (Figure 4-2) appears. Minimum voltage is first; the status display on both
the UUT and the VXI main panel) reads
V_0
. Monitor DVM and click the “+” button to
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......