KLP070312
3-17
(1) To select
NORM
, tap the
DC OUTPUT
switch to save selection; unit returns to
RELY
menu.
(2) To select
V->C
or
C->V
, first set the delay. While
V->C
or
C->V
is in the status dis-
play, the
DC AMPERES
display shows the delay in seconds, from 000.5 to 128.0:
NNND
, where
NNN
is an integer from 0 to 128 and
D
will be either .0 or .5 (000.0 is
not permitted). The delay determines the length of time that the unit must be operat-
ing in the new mode before energizing the relay.
The factory default delay time is set to 1.0 seconds to prevent false tripping of the
relay during application of output power to low impedance or highly reactive loads. If
false tripping occurs, this delay may be increased as needed.
To change the delay, press in and rotate either the
VOLTAGE
or
CURRENT
control;
release the control when the
DC AMPERES
display shows the desired delay. Tap
the
DC OUTPUT
switch to save selection; unit returns to
RELY
menu.
NOTE:
Once the relay turns on in
V->C
or
C->V
mode, the relay will go off again when the
unit transitions back to the original mode, or when the output is turned off.
b. To select
MAN
, tap the
DC OUTPUT
switch to save selection. The
DC VOLTS
display
flashes the present relay state:
ON
or
OFF
. To change the state of the relay rotate either
VOLTAGE
or
CURRENT
control until the desired relay state is displayed, then tap the
DC OUTPUT
switch to immediately set the relay to the selected state and return to the
RELY
menu.
c. To select
LIST
, tap the
DC OUTPUT
switch to save selection; unit returns to the
RELY
menu
3. Rotate either
VOLTAGE
or
CURRENT
control until status display shows
RTN
and press
DC
OUTPUT
switch to exit
UTIL
menu, then rotate either
VOLTAGE
or
CURRENT
control until
status display shows
EXIT
and press
DC OUTPUT
switch to exit FUNCTION control.
NOTE: If a SYSTem:SECurity:IMMediate command is issued, the above relay control settings
are reset to the factory default:
FLT - NORM
.
3.2.16.2 QUICK BOOT
Enter
UTIL
menu (see PAR. 3.2.16, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
QkBT
- Quick Boot. If Quick boot is ON, then the next
time the unit is powered up it will go directly to setpoint mode (status display reads
SET
, see
PAR. 3.2.9) without displaying
MODL
,
VIRT
, and
PROT
settings.
• The
DC VOLTS
display indicates whether quick boot is
ON
or
OFF
.
1. Tap the
DC OUTPUT
switch to change the setting. The
DC VOLTS
display begins to blink.
2. Rotate either
VOLTAGE
or
CURRENT
control until the desired quick boot state is displayed,
then tap the
DC OUTPUT
switch to save the setting; the blinking stops.
3. To exit, rotate either
VOLTAGE
or
CURRENT
control until the status display reads
RTN
, then
tap the
DC OUTPUT
switch. To exit the
UTIL
menu rotate either
VOLTAGE
or
CURRENT
control until the status display reads
EXIT
, then tap the
DC OUTPUT
switch.
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......