3-16
KLP070312
•
SYST
- System Configuration (see PAR. 3.2.16.3)
•
INVC
- Analog Input Voltage and Current A to D calibration voltages (see PAR. 3.2.16.4)
•
LBT
- [Standard models only] Loop Back Test used to isolate RS 232 communication
problems (see KLP Developer’s Guide).
•
CAL
- Display last calibration date and/or begin Calibration (see PAR. 3.2.16.6)
•
RTN
- Exit any submenu and return to previous menu
3.2.16.1 INTERNAL RELAY CONTROL
The internal relay contacts available at pins 6, 7 and 8 of the Analog I/O connector can be con-
figured from the front panel to operate in either Fault, Manual or List mode. If FAULT mode (the
factory default) is selected, the relay energizes automatically upon detection of a user-selected
condition. The selections are a)
NORM
(normal), a system fault (see PAR. 1.5.9), b)
V->C
, when
the unit transitions from Constant Voltage (CV) to Constant Current (CC) mode, or c)
C->V
,
when the unit transitions from CC to CV mode. If
V->C
(or
C->V
)
is selected, the amount of
time the unit must be in CC (or CV) mode before the relay energizes may be programmed from
the front panel (see step 2a) or using the SCPI command
OUTP:PROT:DEL
(see KLP Devel-
oper’s Guide).
1. Enter
UTIL
menu (see PAR. 3.2.16, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
RELY
. The
DC VOLTS
display indicates whether the
relay is currently
ON
(energized) or
OFF
(deenergized). Tap
DC OUTPUT
switch to enter
Relay Control menu.
• Rotate either
VOLTAGE
or
CURRENT
control to see the available relay control modes
in the status display. The modes are:
FLT
(factory default) - Relay is set to energize automatically upon detection of one of
three preselected conditions (tap
DC OUTPUT
to access FLT submenu):
NORM
: (factory default) relay energizes immediately upon detection of a system
error:
V->C
: relay energizes immediately upon detection of a system error and, after
a user-programmable delay, the relay energizes upon a transition
from constant voltage (CV) to constant current (CC) mode.
C->V
: relay energizes immediately upon detection of a system error and, after
a user-programmable delay, the relay energizes upon a transition
from constant current (CC) to constant voltage (CV) mode.
RTN
:
Return to
FLT
menu.
MAN -
Relay on/off is controlled manually, directly from the front panel (see below).
Relay state is not retained; relay will be off when unit is turned on
LIST
- Relay on/off is controlled remotely using LIST commands via GPIB or RS 232
interface (see KLP Developer’s Guide).
RTN -
Return to
RELY
menu.
2. To select mode rotate either
VOLTAGE
or
CURRENT
control until status display reads
FLT
(factory),
MAN
(manual), or
LIST
. Then press
DC OUTPUT
switch to save selection.
a. If
FLT
is selected, rotate
VOLTAGE
or
CURRENT
control to select
NORM
,
V->C
, or
C->V
. in the status display.
Содержание KLP 10-150-2
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......
Страница 78: ......
Страница 90: ......