Troubleshooting 4-9
A/D converter
The SourceMeter unit uses a multi-slope charge balance A/D converter with a single-slope
run-down. The converter is controlled by gate array U610. Commands are issued by the MPU
on the digital board through communications opto-isolators to U610, and U610 sends A/D read-
ing data back through opto-isolators to the digital board for calibration and processing.
Active guard
The Model 2430 has an active guard or “six-wire ohms” circuit used to measure complex de-
vices. This circuitry provides a low-current (50mA) equivalent of the voltage on output HI. If
the unit is in the SV mode, the low-current equivalent of the source voltage will appear on the
guard terminal. If the unit is in the SI mode, the voltage on output HI is equal to the source cur-
rent multiplied by the external resistance value. An equivalent voltage will be generated by the
guard circuit, and a guard sense terminal is provided to sense around the voltage drop in the
guard leads since significant current can flow (50mA).
Digital circuitry
Refer to Figure 4-5 for the following discussion on digital circuitry.
The core digital circuitry uses a Motorola 68332 microcontroller running at 16.78MHz. The
memory configuration includes two 256K
×
8-bit flash EEPROMs and two 128K
×
8-bit RAMs
used in parallel to utilize the 16-bit data bus of the MPU. The RAM is battery backed-up,
providing continued storage of data buffer information during power-down cycles, and flash
ROM support allows internal firmware upgrades using either the serial or GPIB port for
downloading new firmware. All calibration constants and the save 0 setup are stored in a
separate serial EEPROM. Setups 1 through 4 are stored in battery backed-up RAM.
External communication is provided via GPIB and serial interfaces. A 9914 GPIB IEEE-488
standard interface IC is used for the GPIB, and a 68332 Queued Serial Module (QSM) provides
the serial UART. For internal communications, the Time Processing Unit (TPU) is used for serial
communications with the front panel display module, and both the TPU and QSM handle digital-
to-analog interfacing.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......