![Keithley SourceMeter 2430 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803052.webp)
2-20
Calibration
Table 2-9
Voltage range calibration commands
Step Command/procedure
1
Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
:SOUR:VOLT:RANGE <Range>
:SOUR:VOLT -<Source_value>
Take DMM reading.
2
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
3
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:VOLT 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:VOLT +<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:VOLT 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0V.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0V.
Read actual output value.
Calibrate source positive zero.
1. Perform complete procedure for each range, where <Range> = 0.2, 2, 20, and 100, and<Source_value> = 0.2, 2,
20, and 100.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
3.
Calibrate the 10
μ
A to 1A current ranges using the procedure summarized in Table 2-11.
For each range:
•
Send the :SOUR:CURR:RANG command to select the source and sense range
being calibrated. For example, for the 1mA range, the command is:
:SOUR:CURR:RANG 1E-3
•
Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
•
Note and record the multimeter reading.
•
Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 1mA range
would be:
:CAL:PROT:SOUR -1.025E-3
:CAL:PROT:SENS -1.025E-3
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......