Calibration Programs C-3
4.
Ensure the digital multimeter primary address is set to its factory default setting (22).
5.
Ensure that the computer bus driver software (CECHP.EXE) is properly initialized.
6.
Enter the QBasic editor, and type in the program below.
NOTE
Be sure to change the OneOhm parameter to the characterized value of 1
Ω
resis-
tor.(See Section 2 for details.)
7.
Check thoroughly for errors, then save it using a convenient filename.
8.
Run the program, and follow the prompts on the screen to perform calibration. For test
connections, refer to the following figures in Section 2:
•
DMM volts connections: Figure 2-1.
•
10
μ
A to 1A range current connections: Figure 2-2.
•
1
Ω
resistor and DMM connections for 3A range: Figure 2-3.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......