2-12
Calibration
10.
Press ENTER. The Model 2430 will source 0mA and at the same time display the fol-
lowing:
DMM RDG: +00.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
11.
Note and record the DMM reading, and then adjust the Model 2430 display to agree with
the actual DMM reading. Note that the display value adjustment limits are within ±1%
of the present range.
12.
After adjusting the display value to agree with the DMM reading, press ENTER. The
unit will then display the following:
I-CAL
Press ENTER to Output -10.000
μ
A
13.
Press ENTER. The Model 2430 will source -10
μ
A and display the following:
DMM RDG: -10.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
14.
Note and record the DMM reading, and then adjust the Model 2430 display to agree with
the DMM reading. Again, the maximum display adjustment is within ±10% of the
present range.
15.
After adjusting the display value to agree with the DMM reading, press ENTER and note
that the instrument displays:
I-CAL
Press ENTER to 00.000
μ
A
16.
Press ENTER The Model 2430 will source 0
μ
A and simultaneously display the follow-
ing:
DMM RDG: +00.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
17.
Note and record the DMM reading, and then adjust the display to agree with the DMM
reading. Once again, the maximum adjustment is within ±1% of the present range.
18.
After adjusting the display to agree with the DMM reading, press ENTER to complete
calibration of the present range.
19.
Press EXIT to return to the normal display, and then select the 100
μ
A source range using
the EDIT and up RANGE keys. Repeat steps 2 through 18 for the 100
μ
A range.
20.
After calibrating the 100
μ
A range, repeat the entire procedure for the 1mA through 1A
ranges using Table 2-4 as a guide. Be sure to select the appropriate source range with the
EDIT and up RANGE keys before calibrating each range.
21.
After calibrating the 1A range, connect the 1
Ω
characterized resistor and DMM to the
Model 2430 INPUT/OUTPUT jacks, as shown in Figure 2-3.
22.
Select the DMM DC volts function.
23.
Repeat steps 2 through 18 for the 3A range using Table 2-4 as a guide. When entering
the DMM reading, use the calculated current as follows: I = V/R, where V is the DMM
voltage reading, and R is the characterized value of the 1
Ω
resistor.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......