Calibration 2-3
•
If an error occurs during calibration, the Model 2430 will generate an appropriate error
message. See Appendix B for more information.
WARNING
The maximum common-mode voltage (voltage between LO and chassis
ground) is 250V peak. Exceeding this value may cause a breakdown in in-
sulation, creating a shock hazard.
CAUTION
The maximum voltage between INPUT/OUTPUT HI and LO or 4-WIRE
SENSE HI and LO is 125V peak. The maximum voltage between INPUT/
OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO
and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result
in instrument damage.
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
NOTE
Calibration constants are stored in volatile memories of the Model 2430, which
are protected by a replaceable battery when power is off. Typical life for the bat-
tery is approximately 10 years, but the battery should be replaced if the voltage
drops below 2.5V regardless of age. See Section 4 for battery replacement de-
tails.
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use al-
ternate equipment as long as that equipment has specifications comparable to those listed in the
table. For optimum calibration accuracy, test equipment specifications should be at least four
times better than corresponding Model 2430 specifications.
Table 2-1
Recommended calibration equipment
Description
Manufacturer/Model
Specifications
Digital Multimeter
Hewlett Packard
HP3458A
DC Voltage
1
DC Current
1
1V:
10V:
100V:
10
μ
A:
100
μ
A:
1mA:
10mA:
100mA:
1A:
±5.6ppm
±4.3ppm
±6.3ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
Precision Resistor
2
Isotec RUG-Z-1R00-0.1
1
Ω
, ±0.1%, 100W
1. 90-day, full-range accuracy specifications of ranges required for various measurement points.
2. Necessary for calibration of 3A current range. Resistor must be characterized to ±300ppm or better using recom-
mended DMM before calibrating 3A range.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......