Calibration 2-23
NOTE
Calibration will be temporary unless you send the SAVE command. Also, cali-
bration data will not be saved if (1) calibration is locked, (2) invalid data exists,
or (3) all steps were not completed.
Step 6: Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the entire
instrument is properly calibrated. In some instances, however, you may want to calibrate only
certain ranges. To do so, complete the entire procedure only for the range(s) to be calibrated.
Keep in mind, however, that you must complete all parameter steps for each source or sense
range. Also be sure to set calibration dates and save calibration after calibrating the desired
range(s).
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......