![Keithley SourceMeter 2430 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803028.webp)
Digital Multimeter
2430 1KW PULSE SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
125V
PEAK
125V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
4- WIRE
SENSE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
!
®
EXIT
ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
2
3
0
1
6
7
8
9
4
+/-
5
STORE RECALL
EDIT
Model 2430
Input HI
Input LO
1
Ω
Resistor
Figure 1-3
Connections for
3A range current
verification tests
1-12
Performance Verification
4.
Verify output current accuracy for the 3A range. Be sure to:
•
Select the 3A source range.
•
Set the Model 2430 output current to the correct 3A output value.
•
Verify that the multimeter reading is within the 3A range limits given in Table 1-5.
(Since the value of the 1
Ω
resistor value is assumed to be the same as its nominal
value, the DMM voltage reading is the same as the sourced current.)
5.
Repeat the procedure for a negative 3A current output value.
6.
Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2430
source range
Model 2430
output current setting
Output current limits
(1 year, 18°C to 28°C)
10mA
100mA
1mA
10mA
100mA
1A
3A
1
10.0000mA
100.000mA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
9.9947 to 10.0053mA
99.949 to 100.051mA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
2.99543 to 3.00457A
1. See separate procedure for 3A range. DMM voltage reading is same as sourced current.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......