Performance Verification 1-3
Also, allow the test equipment to warm up for the minimum time specified by the manufactur-
er.
Line power
The Model 2430 requires a line voltage of 100 to 240V and a line frequency of 50 or 60Hz.
Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equipment
as long as that equipment has specifications comparable to those listed in Table 1-1. Keep in mind,
however, that test equipment uncertainty will add to the uncertainty of each measurement. Gener-
ally, test equipment uncertainty should be at least four times better than corresponding Model 2430
specifications. Table 1-1 lists the specifications of the recommended test equipment, including
maximum allowable uncertainty for alternate test equipment (shown in parentheses).
Table 1-1
Recommended verification equipment
Description
Manufacturer/Model
Specifications
Digital Multimeter
Hewlett Packard
HP3458A
DC Voltage
1
1V:
10V:
100V:
±5.6ppm
±4.3ppm
±6.3ppm
DC Current
2
10
μ
A:
100
μ
A:
1mA:
10mA:
100mA:
1A:
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
Resistance Calibrator
Fluke 5450A
Resistance
2
1.9
Ω
:
19
Ω
:
190
Ω
:
1.9k
Ω
:
19k
Ω
:
190k
Ω
:
1.9M
Ω
:
19M
Ω
:
±65ppm (±460ppm)
±23ppm (±280ppm)
±10.5ppm (±230ppm)
±8ppm (±200ppm)
±7.5ppm (±195ppm)
±8.5ppm (±200ppm)
±11.5ppm (±180ppm)
±30ppm (±635ppm)
Precision Resistor
3
Isotec RUG-Z-1R00-0.1
1
Ω
, ±0.1%, 100W
1. 90-day, full-range accuracy specifications of ranges required for various measurement points.
2. 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alter-
nate test equipment shown in parentheses.
3. Required for verification of 3A current range. Characterize resistor to ±300ppm or better using recommended DMM
before verifying 3A current measurement range.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......