Command Reference B-9
Error 505
— Will occur when a :CAL:PROT:SOUR command generates an invalid source off-
set calibration constant.
Error 506
— Will occur when a :CAL:PROT:SOUR command generates an invalid source gain
calibration constant.
Error 50
7 — Will occur when a :CAL:PROT:SENS command generates an invalid measure-
ment offset calibration constant.
Error 508
— Will occur when a :CAL:PROT:SENS command generates an invalid measure-
ment gain calibration constant.
Table B-4
Calibration errors
Error
number Error message
+500
“Date of calibration not set”
+501
“Next date of calibration not set”
+502
“Calibration data invalid”
+503
“DAC calibration overflow”
+504
“DAC calibration underflow”
+505
“Source offset data invalid”
+506
“Source gain data invalid”
+507
“Measurement offset data invalid”
+508
“Measurement gain data invalid”
Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it
is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query
to read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an er-
ror occurs, send the following command: *SRE 4. This command will enable SRQ when the
EAV bit is set. You can then read the status byte and error queue as outlined above to check for
errors, and to determine the exact nature of the error.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......